Machine Recycling Discussion and Moderated Newsgroup   >   Full Machines   >   Incomplete Machines   >   South   >   New   >   B&l storz millennium phaco vitrectomy a/p CX3173

B&l storz millennium phaco vitrectomy a/p CX3173


MILLENNIUM ELITE ANTERIOR/POSTERIOR SYSTEM - CX3173.
Venturi vacuum, Bipolar Cautery, Air/Scissors, Dual Illumination, Viscous Fluid, High Speed Vitrectomy System, plus IRIDEX Green Laser.
One year warranty available at additional charge.
Complete with Cart IV POLE, One phaco handpiece, one High Speed Cutter, Silicone Oil injection, Footpedal, One Frag Handpiece.
I have two complete systems ready to go.
Lightning High Speed Vitrectomy Module
With the Millennium Microsurgical System's Lightning High Speed Vitrectomy Module, the vitrectomy procedure is performed with an electrically operated guillotine type cutter that draws the vitreous material into the port. The vitreous is then cut and aspirated into the disposable collection container through the attached tubing.
Phacoemulsification and phacofragmentation refer to the process of ultrasonic disintegration of the lens using a vibrating needle operating at a frequency above the audio range. Phacoemulsification refers to an anterior chamber procedure, and phacofragmentation refers to a posterior chamber procedure. The Millennium Microsurgical System Ultrasound Module is available as a single port phacoemulsification module, a single port phacofragmentation module, or a dual port module which supports both (Elite). The electrical connections are identified by a white ring for phacoemulsification and a yellow ring for phacofragmentation.
Module provides the capability to generate the required injection pressures and aspiration vacuums for injecting viscous fluids and basic salt solution into the eye, as well as aspirate viscous fluids from the eye during posterior segment surgery. The Viscous Fluid Module will deliver up to 70 psi (482.6 kPa, 4.8 bar) of pneumatic pressure to the disposable tube set. All other posterior system functions except phacofragmentation may be operated simultaneously or in conjunction with this module...
Fiber optic illumination for intraocular viewing is accomplished with the Millennium Microsurgical System's Illumination Module. Two ports provide simultaneous operation with independent control. Each source may be independently adjusted for intensity. The light transmitted to the patient will have a spectral content between 400 nm and 750 nm, and an average luminous output of 60 candela per square millimeter. The Function Switch may be programmed to turn lamp 1 or lamp 2 On or Off. The Lamp window must be displayed for the rocker switch to have this control.
The Laser module is controlled by a separate foot switch that is integrated into the Millennium system. The convenience of all of the systems contained in one cart makes the set up and operation of the Millennium that much easier. Comes with 6 goggles and a variety of fibers to get you started.
Bipolar coagulation is accomplished with the Millennium Microsurgical System's Coagulation Module. Bipolar forceps or pencil handpieces are used as electrodes.
Coagulation is selectable at all times, and available when displayed on the screen. Coagulation power may be adjusted from 0% to 100% of the output power using the up/down arrow keys as follows:
Between 0% and 20%, changes are in increments of 1% with each key press.
Between 20% and 40%, changes are in increments of 2% with each key press.
Between 40% and 100%, changes are in increments of 5% with each key press.
Universal Power Input (100 - 240 VAC, 50/60 Hz).
Compressed air or nitrogen - 80 to 100 pounds per square inch
(5.5 to 6.9 bar). Venturi aspiration system only.
21 in. (H) x 21 in. (W) x 20 in. (D) (53 cm x 53 cm x 51 cm)
Cart: 35 in. (H) x 26 in. (W) x 21 in. (D) (89 cm x 66 cm x 53 cm)
Weight: Computer unit and base unit with all module bays filled:
Cart with foot controller stowed: 110 Pounds (50 kg)
CE Marked to IEC 601-1, ETL Listed to UL 2601-1, CSA Listed to C22.2 No. 601.1-M90.
Bottle Infusion/Air Fluid Exchange:
Flow up to 5 standard cubic feet per hour.
Bipolar Coagulatar: 7.5 Watts maximum 100 ohms.
PhacoemuIsification/Phacofragmentation:
Frequency: 28.5 kHz. Power: 35 Watts.
Fixed, Linear, Dual-Linear and pulsed modes.
Range: 30 to 750 cuts per minute.
Range: 30 to 300 cuts per minute.
Fixed, linear, single cut and proportional closure modes.
PhacoemuIsification: 15 mmHg to 250 mmHg
Fragmentation: 15 mmHg to 400 mmHg
Irrigation / Aspiration: 0 mmHg to 550 mmHg
Vitrectomy: 0 mmHg to 550 mmHg
Dual Outputs: 6,000 candela per square centimeter.
Lamp life: 500 hours minimum.
Travel: 100 cm (39.4 in.), pole integral to cart.
Flat panel, liquid crystal, full coIor.
MEDiCAL SURGiCAL EQUiPMENT SOLUTiONS



B&l storz millennium phaco vitrectomy a/p CX3173